The tester has been improved so that the operator can conduct the test easily using the jog dial type controller and touch panel display, which is equipped with voice guidance. With these functions, the operator can concentrate on observing the condition of the test specimen, while controlling the burner to adjust the distance between the flame and the test specimen.
UL-94, JIS-K6911,ASTM-0635, 03801,04804,05025,05207,150-1210
Product Description
- No.252-UL-94 PLASTIC FLAMMABILITY TESTER (ELECTRIC SYSTEM)
- The tester has been improved so that the operator can conduct the test easily using the jog dial type controller and touch panel display, which is equipped with voice guidance. With these functions, the operator can concentrate on observing the condition of the test specimen, while controlling the burner to adjust the distance between the flame and the test specimen.
- No.252-UL-94 PLASTIC FLAMMABILITY TESTER (MANUAL)
- This tester is used to evaluate the burn-resistance of plastic according to the vertical flaming test (VB) and horizontal flaming test (HB). By adding the option jigs, the tester can also carry out other test methods (5V, VTM, HBF, HF).
Specification
Model | No.252-UL-94 (Manual) | No.252-UL-94 (Electric System) |
Chamber | 1,000 x 555 x 905 mm (VolumeApprox. 0.5 m3)l1,000 x 555 x 905 mm (VolumeApprox. 0.5 m3) |
Specimen | L125 ± 5 mm, W13.0 ± 0.5 mm, T13 mm or less |
Burner | L1oo ± 1o mm, Inner Diameter cp9.5 ± 0.3 mm |
Burner Base | Slide type (With Operation Handle), | Iide type (With Jog Dial Type Controller), |
Inclination Angle 45 ± 2″, 20 ± 5″ (5 V) Inclination Angle 45 ± 2″, 20 ± 5″ (5 V) | |
Flow Meter: Max.150 ml/min (Scale 2 mVmin) (Standard 105 ml/min) Flow Meter: Max. 1,000 ml/min (Scale 10 mVmin) (Standard 965 ml/min) Manometer: Max. 200 (± 100) mmH20 (Scale 1 mmH20) (Standard 10 mmH20) Timer, Built-In Thermometer | |
Burner Calibration | |
Device | |
Accessories | Flame Gauge (20 mm) (Option: 125 mm & 38 mm), Ring Stand, Wire Mesh, Dry Cotton Pad, Fireproof Gloves, Gas Lighter |
Horizontal Burning Support Fixture (HB), Plaque Support Fixture (5 V), | |
Option | Spring Clamp (VTM), Foam Support Fixture |
Power Source | AC 100 V, 1-Phase, 15A, 50/60Hz | AC 100 V, 1-Phase, 10 A, 50/60 Hz |
Heat Source | Methane Gas (Purity: Over 98 %) |
Dimensions/ Weight | W1,510 x D610 x H1,100 mm/200 kg lw1,500 x D615 x H1,100 mm/230 kg |
(Approx.) |